SD-H3000A++
Electromagnetic compatibility(EMC):
Radiation: Conform IEC 60601-1-2
Anti-interference performance: Conform IEC 60601-1-2
Critical-care ventilator:
General instruction: patient device: length 455mm x width 400mm x height 300mm
weight would be appximately 12.5kg.
Triggering: flow rate trigger and pressure trigger.
Gas supply:
the supplied gas must be fresh gas which is water-free, oil-free or no impurity particles, and its contents must be below the following standards:
Air: H2O < 79mg / m3, oil < 0.5 mg / m3
Oxygen: H2O < 20mg / m3
Inlet pressure: 0.28~0.6 MPa
Connecting standard NIST or DISS standard could be used.
Patient system gas transporting connector:
Cone port: male connector 22mm;
female connector 15mm.Conform ISO 5356-1:2008.
Exhaust port: male connector 22mm Cone
Suitable patient arrange:
Infant(≥3Kg), child, adult patient.
Operation condition:
Operating temperature range: 10℃ ~40℃.
Relative humidity: 5%~95%, Non-condensing.
Atmosphere pressure: 500 - 1060hPa
Altitude: ≤5000m
Non-operating status condition:
Anti-collision: peak acceleration: 15g
Pulse duration: 6 ms
The number of collisions: 1000
Temperature: from -20℃ to +60℃
Relative humidity ≤95 %
Atmosphere pressure: 500 - l060hPa
Altitude: ≤5000m
Testing mode:
The flowrate measurement of this ventilator used ATP manner, that is: The gas flowrate and flow quantity were transformed into the volume flow quantity under the current environment (current atmosphere pressure, current temperature and current humidity).
Inspiratory channel:
Pressure decreasing:
When flowrate was 60L/min, the maximum would be 6cmH2O.
When flowrate was 30L/min, the maximum would be 3cmH2O.
When flowrate was 5L/min, the maximum would be 1cmH2O.
Inspiratory resistance: 6cmH2O @ 60L/min
Gas transportation: Flowrate arrange:0~180L/Min.
Expiratory channel:
Pressure decreasing:
When flow rate would be 60L/Min, the maximum is less than 2cmH2O.
When flow rate would be 30L/Min, the maximum is less than 1cmH2O.
When flow rate was 5L/min, the maximum is less than 0.5cmH2O.
Expiratory resistance: 3cmH2O @ 60L/min
Expiratory flow rate measuring: Flow rate arrange: 0~180L/Min.
Breathing system:
VBS compliance: <4mL/cmH2O
Recommended breathing system compliance: <50mL/cmH2O
Recommended breathing system Resistance: <3cmH2O @ 60L/min
Recommended filter
Resistance: <3cmH2O @ 60L/min; Deadspace: <30mL
Device noise: <65dB
About NIV mode:
Depending on the device configuration, it can be used for noninvasive ventilation when the device is configured in a noninvasive mode. See 7.2.2.2 for instructions.